MAIN UNIT |
Channel Configuration | Single and dual channel |
Moisture Content | ppmW |
Temperature | °C and °F |
Water Dew Point | °C and °F |
Sample Flow Rate | Status indication: Low-Good-High (<0.1->0.3 l/min) |
Sensor Technology | Michell Ceramic Moisture Sensor |
Temperature Measurement | Thermistor-accuracy ±0.2°C |
Operating Pressure | 50 Barg (max) |
Calibration Range | -100°C dp to +20°C dp |
Measurement Range | -120°C dp to +30°C dp (0.001 ppmW to saturation) |
Sample Flow | Up to 1 l/min (0.1-0.2) recommended |
Accuracy | Dew point: ±1°C between –59.9 and +20°C dp Moisture content: ±10% of reading Dew point: ±2°C between –60 and -100°C dp Moisture content: ±20% of reading Analysis Pressure: ±0.25% FS |
Resolution | 0.1°C between +20°C dp and -80°C dp 1°C between -80°C dp and -100°C dp |
Temperature Coefficient | Algorithm compensation |
HMI |
Keypad Interface | Capacitive touch-screen through glass |
Display | Vacuum flourescent |
Data Logging | Two non-isolated 4-20 mA per measurement channel. 500 Ω maximum load. Parameter and range selectable by the user Communications and Modbus: RTU @ 9600 baud-rate output Alarms: Two volt-free contacts per channel; one for process value Alarm status also available via Modbus communications |
ENCLOSURE |
Type | Flameproof EExd |
Construction | Cast copper free aluminium (IP66) |
Finish | Chromate primer, Polyester P9010 powder coated (black). Meets BS3900 |
GENERAL |
Sample Connections | 1/4” NPT (female) |
Weight | 22.5kg |
Operating Supply Voltage | 85 to 264 VAC, 47 to 63 Hz, 110 to 300 VDC, 180 W |
Operating Environment | -20°C to +60°C up to 95% RH, Indoor or Outdoor, shaded location |
REMOTE INTERFACE |
Indoor Safe Area Installation | Up to 31 Michell EExd Process Instrumentation Range analyzers in any combination of models via Modbus (Optional) |
Host Capacity | Modbus |
Display | Backlit color 15cm / 6” LCD |
User Interface | Integral mouse, on-screen keyboard |
Function | View and modify analyzer operating settings. Record parameters to file and view virtual chart recorder and Analyzer diagnostics |
Ports | LAN, USB2 |
Data Download | CSV (comma separated variable) format for easy import to Excel™ |
Remote Access | OPC (OLE for Process Control) over a LAN Imbedded web page/unique IP address for PC access |
Power Supply | 100-280 V, 50/60 Hz |
HAZARDOUS AREA CERTIFICATION
ATEX Certification: II 2 G Code: EEx d IIB + H2 T4 (135°C) IP66 Tamb -40 to + 60°C Certificate number: TRL06ATEX21093X CCSAUS approved: Div 1, Class 1, Group B, C, D Certificate number: 1722693 |